Changsha Panran Technology Co., Ltd.
How to calibrate a 400bar pressure gauge and what instruments are required
Source: | Author:SW | Published time: 2025-02-07 | 34 Views | Share:

To calibrate a 400 bar pressure gauge, you need to ensure you have the proper equipment and follow a systematic process. Here's how to do it:

Required Equipment:

  1. Precision Pressure Source (Calibrator):

    • A high-accuracy pressure calibrator or pressure pump that can generate and control pressure up to 400 bar (e.g., a pneumatic or hydraulic pump). The pressure source should be capable of applying pressures from 0 to 400 bar with minimal deviation.

  2. Reference Pressure Gauge or Deadweight Tester:

    • A highly accurate, traceable reference pressure gauge or a deadweight tester to compare the readings of the pressure gauge being calibrated. The reference gauge must have an accuracy that meets or exceeds the specification of the pressure gauge being calibrated (usually ±0.25% to ±0.5% of full scale for industrial calibrations).

  3. Digital Multimeter (Optional):

    • If the pressure gauge is a transmitter or has an electronic output (e.g., 4-20 mA), you will also need a digital multimeter to measure the output signal.

  4. Connecting Fittings and Tubing:

    • Proper fittings, adapters, and tubing to securely connect the pressure gauge to the pressure source and reference gauge without leaks.

  5. Calibration Software or Handheld Communicator (Optional for Digital Pressure Gauges):

    • For digital pressure gauges or those with HART or other communication protocols, a handheld communicator or calibration software might be required to access and adjust calibration settings.


Step-by-Step Calibration Process:

1. Prepare the Setup:

  • Connect the Pressure Gauge:
    Connect the 400 bar pressure gauge to the pressure calibrator and the reference pressure gauge using the appropriate fittings and tubing. Ensure that there are no leaks in the setup.

  • Set Zero Pressure:
    Ensure both the pressure gauge and the reference gauge read zero when no pressure is applied. If necessary, adjust the zero of the gauge under calibration.

2. Apply the Pressure:

  • Start with Low Pressure:
    Begin by applying a low pressure, such as 0 bar (atmospheric pressure) or a low value, and check that both gauges (the one being calibrated and the reference gauge) are showing the same reading.

  • Increase Pressure in Steps:
    Gradually increase the pressure in known increments (e.g., 10%, 25%, 50%, 75%, 100% of the full scale). For a 400 bar gauge, this would mean applying pressures like 40 bar, 100 bar, 200 bar, 300 bar, and 400 bar.

3. Check and Adjust Readings:

  • Compare Readings:
    At each pressure increment, compare the reading on the pressure gauge with the reference pressure gauge. The difference should be within the specified tolerance for the instrument.

  • Adjust Calibration (if necessary):
    If the readings on the gauge being calibrated are outside the acceptable tolerance compared to the reference, make adjustments. Some analog pressure gauges have mechanical adjustments for zero and span, while digital ones may require recalibration via a communicator or software.

4. Check the Full Range:

  • Test at Full Scale:
    At 400 bar, ensure the gauge is displaying the correct pressure. This is the critical point to verify that the gauge can handle the maximum rated pressure.

  • Intermediate Points:
    Repeat the check at intermediate pressure values (e.g., 50 bar, 150 bar, 250 bar) to ensure the gauge is linear across the entire range.

5. Document the Calibration:

  • Record the pressure values and corresponding readings from both the calibrated gauge and the reference gauge for future reference and certification. If the instrument was adjusted during the process, document the changes made.

6. Verify Calibration:

  • After completing the calibration process, it’s a good practice to recheck the gauge by applying pressures in the same increments to confirm that the adjustments have been effective and the gauge is accurate across its full range.

7. Final Step – Reinstall the Gauge:

  • After calibration is complete, carefully reinstall the calibrated pressure gauge into its operating environment and ensure all connections are secure.


Important Notes:

  • Traceability: Ensure that the reference pressure gauge or calibrator is calibrated with traceability to national or international standards (e.g., NIST or UKAS). This ensures the accuracy of the calibration results.

  • Pressure Medium: If calibrating a hydraulic pressure gauge, be sure the pressure source uses the same medium (e.g., hydraulic oil) as the gauge is designed for.

  • Safety Precautions: Always handle high-pressure equipment carefully and wear appropriate personal protective equipment (PPE). Ensure that the calibration process is performed in a safe environment.

Would you like more specific details on any particular part of the process or equipment recommendations?