How to test a pressure switch?
Source: | Author:J | Published time: 2025-05-23 | 6 Views | Share:

A pressure switch can be tested if functioning properly by supplying pressure until the switch is triggered once the set pressure is reached. The pressure switch has a scale to see the settings of the required pressure. Once the switch is triggered, it will start or stop a process.

During calibration, a known pressure source from a calibrator is used in order to determine the exact value of pressure when the switch is triggered.  Just compare the set pressure value and the pressure reading from the calibrator.


The analog or mechanical type of pressure switch can be tested or calibrated even it is removed from its power source. We will be using a multimeter on its OHM meter function or to the most advanced type is the Open and Short functions where it emits a sound when the circuit is Shorted and stop when it is Opened.

Pressure Switch Calibration Procedure

In this procedure, we will use the Baumer or Bourdon Haenni brand of pressure switches which are used to monitor pressure on a compressor system in a gas industry. This procedure is also applicable to a differential pressure switch type. Moreover,  this method can also be used to most pressure switches as long as proper fluid and fittings used are compatible including the digital pressure switches.

Objective:

To define the calibration procedures for Pressure Switch using Electronic Deadweight Tester (EDWT), pressure module with a pneumatic hand pump and a process calibrator.

Calibration Method:

The pressure generated in EDWT or pressure module with the hand pump is directly compared to the set value of the Unit Under Calibration (UUC) which is the pressure switch. The pressure switch triggering mechanism ( OPEN/CLOSE) is monitored or detected through the use of the process calibrator or simply a multimeter (ON and OFF triggering using the tone emitted).

Requirements:

  • Warm-up time ( Unit Under Calibration): At least 1 hour for proper stabilization

  • Temperature: 23 +/- 5 deg C

  • Humidity: 50 +/- 30%

  • Perform 3 trials for each point or range

  • Measurement Data Sheet (MDS)– for the recording of results.

Reference Standards

The reference standards and equipment needed are:

  1.  a Digital Multimeter

  2. Deadweight Tester (EDWT)– for high-pressure source

  3. PANRAN Pressure module with a pneumatic hand pump – for low-pressure source

  4. Thermohygrometer– for temperature and humidity monitoring

  5. Set of fittings

  6. Cleaning materials

Calibration set up

For high-pressure ranges, we will use an Electronic Deadweight Tester ( EDWT) and still a multimeter or Panran 235 for open-close monitoring of the switches.


Calibration Procedure:

  1. Check the pressure switch for any visual defects that can affect its accuracy. Discontinue calibration if any defect is noted.

  2. Observe proper stabilization time of at least 1 hour.

  3. Prepare the measurement data sheet (MDS) and record all necessary details or information (Brand, Model, serial #, etc).

  4. Determine the test point to be calibrated.

  5. Connect the pressure switch to the pressure calibrator as shown in the set up above.

  6. Connect the Panran 235 or a digital multimeter as shown below. Set the meter to its Open/Short function.

7. SET action: Apply pressure from the deadweight tester or hand pump until the switch is triggered. You know that the switch is triggered when a continuous tone from the multimeter or process calibrator is heard.  This is the case for a NO (normally open) setting of the pressure switch.

(For Normally Close (NC) switch, an emitted tone will stop once the switch is triggered)


  1. Wait for the display on the pressure calibrator to stabilize, then get the reading. Record readings on the Measurement Data Sheet (MDS).

  2. RESET action: After recording the value of increasing pressure, slowly decrease the pressure then observe if the emitted tone will stop. Record the indicated pressure value.

    10.Repeat this for 3 times.

    11.Check readings if within the accuracy defined by the manufacturer, for example , accuracy = +/- 1% of Full Scale. If the readings are already within limits update the corresponding record, do labeling and sealing and issue to owner, otherwise, do necessary repair or adjustment.

    12.End of Verification

Pressure switch adjustment techniques

Adjustment technique when using these type of pressure switches ( Baumer / Bourdon Haenni with model  RP2El or RPPNAA3601-0201).