Tools Required
Dry well (e.g., HSIN150B with ±0.05°C stability)10
Standard RTD (e.g., PT100)
Thermal paste
Procedure
Insert Probe: Place RTD ≥150 mm deep into well6.
Set Parameters: On touchscreen, input 0°C and 100°C test points.
Stabilize: Wait 15 minutes for ±0.05°C stability10.
Record Data: Export deviations via USB.
Common Error
Avoid uneven thermal contact—use conforming inserts for odd-shaped probes5.