Changsha Panran Technology Co., Ltd.
Step-by-Step: Calibrating RTDs Using Touchscreen Dry Wells
Source: | Author:Y | Published time: 2025-06-27 | 0 Views | Share:

Tools Required

  • Dry well (e.g., HSIN150B with ±0.05°C stability)10

  • Standard RTD (e.g., PT100)

  • Thermal paste

Procedure

  1. Insert Probe: Place RTD ≥150 mm deep into well6.

  2. Set Parameters: On touchscreen, input 0°C and 100°C test points.

  3. Stabilize: Wait 15 minutes for ±0.05°C stability10.

  4. Record Data: Export deviations via USB.

Common Error

Avoid uneven thermal contact—use conforming inserts for odd-shaped probes5.