Resistance Temperature Detectors (RTDs) are widely used in industrial and laboratory environments due to their high accuracy, stability, and repeatability. However, like any temperature sensor, RTDs require regular calibration to ensure precise measurements over time.
The PR611 Series Multifunctional Dry Block Calibrator is designed to calibrate RTDs with high precision, featuring dual-zone temperature control, built-in electrical measurement channels, and an automatic calibration mode. This article provides a detailed step-by-step guide to calibrating RTDs using the PR611 calibrator.
RTDs work by measuring the change in electrical resistance as temperature changes. However, factors like wire aging, mechanical stress, and environmental conditions can cause RTDs to drift from their original calibration.
Regular calibration ensures:
✔ Accurate temperature readings
✔ Compliance with industry regulations
✔ Reliable performance in process control applications
✔ High-accuracy temperature control with ±0.01°C stability
✔ Supports two-wire, three-wire, and four-wire RTD configurations
✔ Built-in electrical measurement for direct resistance comparison
✔ USB storage for automated data logging
✔ PR611 Dry Block Calibrator
✔ RTD sensor to be calibrated
✔ Reference RTD (optional for external reference mode)
✔ Connection wires and terminals
✔ USB drive (for data recording, if needed)
PR611 supports calibration of multiple RTD types, including:
Pt100, Pt500, Pt1000 (Industrial applications)
Cu50, Cu100 (Copper-based RTDs)
Pt10, Pt200 (Specialized high-precision applications)
Check the manufacturer’s specifications for the correct resistance-temperature relationship.
RTDs can be connected using different wiring configurations:
Connect both leads to the PR611 RTD input terminals.
Note: Two-wire RTDs are susceptible to lead wire resistance errors, reducing accuracy.
Connect the common lead to one terminal and the other two leads to separate terminals.
The PR611 automatically compensates for lead resistance, improving accuracy.
Use two pairs of leads to eliminate wire resistance errors.
The PR611 applies a constant current and measures voltage drop across the RTD.
Power on the PR611 and navigate to the Calibration Interface.
Select RTD Measurement Mode.
✔ Internal Reference Mode – Uses PR611’s built-in precision sensor for calibration.
✔ External Reference Mode – Requires a high-precision reference RTD sensor for improved accuracy.
To set the reference mode:
Navigate to RTD Settings.
Select "Internal" or "External Reference".
In the Task Settings menu, define up to 6 calibration temperatures.
Recommended calibration points:
Low range: -50°C to 0°C
Mid range: 100°C to 300°C
High range: 500°C to 700°C
If needed, enable double-stroke mode for both heating and cooling calibration.
Set a stability threshold (e.g., ±0.01°C over 5 minutes).
Define a delay time (e.g., 2 minutes) before recording data after stabilization.
Set the desired temperature setpoint in PR611.
Wait for the calibrator to reach and stabilize at the setpoint.
Once stable, PR611 will automatically measure the RTD resistance.
The PR611 will display both the reference temperature and the RTD reading.
Compare the measured resistance with the expected R(T) curve for the RTD type.
✔ View real-time measurement values on the PR611 display.
✔ Save data to a USB flash drive in CSV format for documentation and compliance.
Error Type | Possible Cause | Solution |
---|---|---|
Large Offset | RTD drift due to aging | Recalibrate or replace RTD |
Nonlinear Response | Mechanical stress or damage | Inspect sensor and wiring |
Fluctuating Readings | Loose connections | Tighten and secure terminals |
If deviations are found, apply correction factors in the temperature controller or data acquisition system.
✔ Use the correct wiring configuration to minimize resistance errors.
✔ Allow enough stabilization time before taking readings.
✔ Perform multi-point calibrations to check accuracy across the full range.
✔ Ensure proper sensor placement for uniform heating.
✔ Use PR611’s automated mode to simplify repetitive calibrations.
✔ Temperature stability of ±0.01°C, ensuring accurate RTD calibration.
✔ Heats from 23°C to 155°C in 17 minutes.
✔ Cools from 23°C to -30°C in 12 minutes.
✔ Measures RTD resistance with ±0.015% accuracy.
✔ Supports two-wire, three-wire, and four-wire configurations.
✔ Saves calibration data for compliance and documentation.
The PR611 Multifunctional Dry Block Calibrator is an ideal solution for RTD calibration, offering high accuracy, stability, and automated calibration functions. Whether used in industrial plants, research labs, or quality control departments, PR611 ensures reliable and repeatable RTD calibration.
By following the step-by-step calibration process outlined above, users can achieve consistent and traceable RTD calibration results with minimal effort.